CHLOR*TEST™ Surface Salt Test


An easier to use, more accurate field test than anything before.

Surface Salt Test

CHLOR*TEST™* was developed for ease of use and to prevent outside and cross contamination. Offers consistent, safe and reliable testing results for chloride contamination.



 

Technical Data

Technical Data
Measuring Range: 5 - 200 ppm
Sampling Time: 1.5 minutes
Color Change: Pink - White
Detectable Limit: 1 ppm
Storage Condition: In a cool and dark place, not exceeding 77°F (25°C )

Instructions for Use

  1. Pour CHLOR* EXTRACT™ into CHLOR* SLEEVE™.
  2. Peel off protective backing, pinch CHLOR* SLEEVE™ and adhere sleeve to surface to be tested.
  3. Massage extract against surface, then peel from surface.
  4. Insert glass titration tube into sleeve. Approximately 2 minutes later, a simple color change will identify the chloride level in both parts per million and micro grams per centimeter squared. Can be used on vertical, horizontal or overhead surfaces.
You can also insert shoelace into side holes of the box and tie it around waist or belt loops for hands-free testing.

Applications

  • Bridges
  • Aircraft
  • Mining facilities
  • Ships
  • Marine structures
  • Storage tanks
  • Cooling towers
  • Pipe lines
  • Power plants
  • Offshore rigs
  • Petrochemical plants
  • Pulp and Paper mills
  • Steel fabrication
  • Electronics

Ambient Condition Correction

No temperature correction is necessary at the sample solution of 41°F (5°C) to 176°F (80°C)

Interferences
Coexistence of Bromide ion, lodide ion or Cyanide ion respectively with Chloride ion gives higher readings. Sulphide ion produces a brown stain and the coexistence with Chloride ion produces a brown stain in the bottom of the stained layer and gives higher readings. PH value within 3.5 to 11 does not affect the readings. Less than pH 3.5 or more than pH 11 gives higher readings.

Chemical Reaction in the Detector Tube: NaCl + Ag2CrO4 = AgCl

Chlor Test Video

 

Prices

top
Prices subject to change    Prices listed in US dollars for domestic shipments within US    For tax exempt orders please call us at 954-946-9454
Item Number Item Prices
CD-1341 CHLOR*TEST™ Surface Salt Test Kit (5 individual tests)
CD-134150 CHLOR*TEST™ Surface Salt Test Bulk Pack (50 individual tests) 913.00
CD-1342 CHLOR*TEST™ “A” Abrasive Kit (4 individual tests) 78.00
CD-1343 CHLOR*TEST™ “W” Water Kit (5 individual tests) 87.00
CD-1344 CHLOR*TEST™ “C” Concrete Kit (4 individual tests) 112.00
* CHLOR*TEST is a registered trademark of CHLOR*RID International



top

Call HAVE QUESTIONS?
    954.946.9454
Email PREFER TO EMAIL?
    gardner@gardco.com
  FIND PRODUCTS LEARN MORE ORDERING INFO ABOUT US
  By Category Email Registry How to Order Contact Info
  By Industry Trade Shows Request Quote History
  New Products Test Lab Info Order Form FAQ's
  Monthly Product Articles/Links Sales & Clearance Policies




  © 2020 Paul N. Gardner, a BYK Instruments Company | Contact | Policies Facebook  YouTube  Blogger