The CAM-Plus is contact angle meter that provides a patented method ("half angle") for quantifying surface energy relative to other measurements on other surfaces. This information makes it possible for accurate prediction of a surface reaction to coatings, eliminating costly trial-and-error methods. It eliminates the difficulties associated with tangent angle measurement methods.
Features
Surface energy is most evident in how a surface interacts with an applied liquid. Like color and texture, surface energy is a physical property. When paint is applied to a sheet of glass, it will typically "ball up" because it does not stick well to the surface. This is surface energy in action. Conversely, when liquid is applied to a different material and disperses, this is also surface energy.
Technical Attributes
Measuring Method: | Sessile Drop, Half-Angle™, Tangent Line |
Measuring Range: | 0-180° |
Repeatability: | ±2° |
Accuracy: | ±1.5° |
Magnification: | 6x (standard), 12x (optional) |
Sample/Specimen Stage Sizes: | CAM-Micro - 2" x 2" x 1"D (50 x 50 x 25mm D) - (Micro) CAM-Film - 10mil max., 1” strips - (Film) |
DOSING: | 50-75 droplets |
Input Voltage: | 120V/50/60Hz or 240VAC (only) |
Dimensions: | 8” W x 15” L x 13” D (PS separate) CE approved |
Weight: | 30 lbs., including power supply |
Standards
ASTM D20.19, D724, D5946 and TAPPI T458 cm-94
Delivery Content
Includes Syringe assembly, post holder for Cam Syringe Micro and Film Specimen Holder Stage, Dove Prism, 6x Projection Lens and Ionizer, Manufacturers Certificate of Inspection
Accessories