Features
Technical Attributes
Power Supply: | 2 x AA (batteries or rechargeable batteries) |
Battery Life: | One measurement per second up to 60,000 readings. (When device is not in use at min. 50 % battery capacity > 1 year) |
Display | IPS-LCD, 2", color, 350 cd/m2, viewing angle 70° all directions |
Material Housing Parts Hard Fiberglass Reinforced: | PA12 GF30 |
Material Housing Parts Soft: | TPE-U |
Material Sensor Holder (Transparent): | Trogamid® (CX7323) |
Scratch Resistant Glass Display: | H6 |
Thickness of Glass Display: | 2 mm |
Dimensions (H x W x D): | 136 x 63 x 42 mm |
Weight: | 132 g |
IP Protection: | IP65 |
Operating Temperature: | -20 °C to +70 °C |
Storage Temperature: | -20 °C to +70 °C |
QNix® 5500 with Integrated Probe, (Optional): | Fe: 3 mm 5 mm Dual Fe/NFe: 3 mm 5 mm |
QNix® 5500 with Modular Probe, (Optional): | Fe: 1.25 mm 3 mm 5 mm NFe: 3 mm 5 mm Dual Fe/NFe: 1.25 mm / 1.25 mm 3 mm / 3 mm 5 mm / 3 mm 5 mm / 5 mm |
Cleaning Instructions: Device: Before cleaning please switch it off! Use a damp lint free cloth with mild detergents. For soiling due to paint and varnish, please use diluted solvents (ethanol, nitro thinner, petroleum ether). Notice! Do not use acetone on soft rubber parts or display! Display: Use a damp, lint free cloth. If very dirty (water is not enough) use mild detergents.Note! Acetone can damage the display |
More Information
All exchangeable standard probes of the QNix® 5500 can also be used in the new QNix® 9500. QNix® thus ensures the
flexibility and compatibility of the new probes which are interchangeable between both models. The QNix® probes have
been completely redeveloped: The standard probes are small and extremely robust. FE-, NFe- and dual probes with
different measuring ranges are available. The coating thickness gauge QNix® 5500 is
available with a fixed
integrated probe and a modular, exchangeable probe. QNix® 5500 changeable probes:
Status Display with RGB-LED on the Probe: | yes |
Measuring Principle: | magnetic: magnetic field amendment / hall-effect Fe / whirl current NFe |
Standards: | DIN EN ISO 2808, DIN 50981, DIN 50984, ISO 2178, BS 5411 (3 & 11), BS 3900-C5, ASTM B 499, ISO 2360, ASTM D 1400, ASTM D 1186, ASTM D 7091 |
Measuring Range: | depending on the probe |
Measuring Speed (Measurements per Minute): | ≥ 120 |
Accuracy of Measurement on Fe Substrates Related to QNix® Reference Standards: | ± (1 µm + 2 % of the reading) to 2.0 mm ± 3 % of the reading from 2.0 mm |
Accuracy of Measurement on NFe Substrates Related to QNix® Reference Standards: | ± (2 µm + 2 % of the reading) to 2.0 mm ± 3 % of the reading from 2.0 mm |
Smallest Measuring Surface Fe / NFe: | Diameter 14.5 mm |
Smallest Curvature Konvex Fe: | 6 mm - Zero point adjustment on original substrate (accuracy checked up to 250 µm) |
Smallest Curvature Konvex NFe: | 50 mm - Zero point adjustment on original substrate (accuracy checked up to 250 µm) |
Minimum Thickness of Base Metal Fe: | 100 µm - Zero point adjustment on original substrate (accuracy checked up to 250 µm) |
Minimum Thickness of Base Metal NFe: | 50 µm - Zero point adjustment on original substrate (accuracy checked up to 250 µm) |
Material Casing Top:* | Stainless steel V4A (1.4305) |
Material Casing Bottom:* | Trogamid® |
Material Measuring Tip:* | Ruby, polished (Al203) |
Dimensions (Height, Diameter): | 34.5 mm, 14.5 mm |
Weight: | 11.9 g |
IP Protection: | IP65 |
Operating Temperature: | -20 °C to +70 °C |
Storage Temperature: | -20 °C to +70 °C |
Max. Surface Temperature: | 80 °C |
Max. Surface Temperature at 1s Measuring Time every 20s: | 100 °C |
* Clean with a damp cloth; chemically resistant against acetone (except soft rubber parts e.g. keyboards or rubber protective frame), nitro dilution and ethanol |
Delivery Content
QNix® 5500 Integrated Model - Handheld QNix® 5500 gauge, Fe- and NFe-reference plate and two test foils, Test certificate, Operation manual, 2 Mignon batteries 1,5 V (AA) alkaline, Plastic case for transport and storage
Accessories