Certificate:
Features
Orifice Interchange (not applicable to certified cups) - Production tolerance on the Gardco Standard ISO cup, the Gardco/ISO Mini Dip Cup and the orifice, whether it be 3mm, 4mm, or 6mm, is sufficiently close to permit interchange of orifices in any one cup or between cups. For convenience and for best possible results, however, it is recommended that such interchanges be held to a minimum.
The orifices for the cup are machined from stainless steel with the exact conical exterior to match the opening in the base of the cup. Positive orifice identification is assisted by "steps" at the base of the orifice as shown in the cross section drawing. (See "Care of Cup" for orifice removal instructions.)
The following is quoted in part from the rationale of the ASTM for the inclusion in ASTM standards, a reference for the design and use of the ISO viscosity cups: “In the 1984 edition of Recommendations of the Committee of Experts on the Transportation of Dangerous Goods, and in similar documents of the International Civil Aviation Organization (ICAO) and of the International Maritime Organization (IMO), an ISO Flow Time Test (ISO 2431) appears. It is used as one of several criteria used to permit shippers of paint and other viscous materials to transport them as a less hazardous material than that indicated by flash point alone. (Request our Flash Point Tester Literature) The three standards involved are: 1. viscosity (ISO 2431), 2. flash point (ISO 1523/ASTM D3941), and 3. solvent separation test (Now being proposed as an ASTM method.)”.
Technical Attributes
Orifice | Seconds Range | Centistoke Range | Midrange Sensitivity ¹ | Calibration Oil No./ Cenistokes ² |
3mm | 27 → 120 | 4.6 → 52 | .05 | #G 10/19Cs |
4mm | 21 → 120 | 19 → 163 | 1.4 | #G 35/68Cs |
6mm | 17 → 120 | 84 → 823 | 7.1 | #G 100/233Cs |
8mm | 10 → 120 | 187 → 2611 | 22 | #G 350/878Cs |
¹ Stated as centistokes per second of efflux time. ² Centistoke values are nominal - actual values printed on labels |
More Information
ISO Dip Cup
The physical dimensions of the new ISO Dip viscosity cups have been chosen to provide just one half of the efflux time in seconds of the Gardco Standard ISO cups of the same number and orifice. The inverted cone bottom shape of the cup terminating in the removable stainless-steel orifice is retained to insure a sharp end point and best possible comparable accuracy. For the first time a dip cup has been developed primarily for field use which is a direct counterpart of the cup used in the laboratory. Gardco is pleased to have made this contribution and to introduce the ISO Mini Dip Cup series along with the new Gardco Standard ISO series. The Gardco Dip series has all of the advantages listed for the Standard series except it does not have a drip ring at the top of the cup & a handle is attached for support. The Dip cup capacity is one-half that of the Standard cup.
Click HERE for more information on Viscosity Cups.
Standards
Procedure
ISO Dip Cup Instructions
ISO Standard Cup Instructions
Accessories