Download Excel Viscosity Chart HERE
Features
Technical Attributes
Item Number | Cup No. | Seconds Range | Centistoke Range |
13573 | 1 | 31 to 60 | 15 to 78 |
13575 | 2 | 19 to 60 | 39 to 238 |
13577 | 3 | 11 to 60 | 63 to 604 |
13579 | 4 | 10 to 60 | 97 to 899 |
13581 | 5 | 10 to 60 | 219 to 1627 |
Certificate of Calibration available at additional charge for all cups. |
Cup No. | Efflux Time Seconds “T” from Centistokes “V” | Centistokes “V” from Efflux Time in seconds “T” |
1 | T = (V + √ (V2 + 6805) ) ÷ 3.18 | V = 1.59T - 1070 ÷ T |
2 | T = (V + √ (V2 +12707) ) ÷ 8.36 | V = 4.18T - 760 ÷ T |
3 | T = (V + √ (V2 + 23529) ) ÷ 20.46 | V = 10.23T - 575 ÷ T |
4 | T = (V + √ (V2 + 32983) ) ÷ 30.26 | V = 15.13T - 545 ÷ T |
5 | T = (V + √ (V2 + 58903) ) ÷ 54.54 | V = 27.27T - 540 ÷ T |
Note: Furnished with each cup is a conversion table for each tenth of a second within the cup range. |
Catalog Number | Oil Number | Use With Cup Number | Nominal Centistokes* at 25°C., 77°F. |
13642 | G-20 | 1 | 35 |
13646 | G-60 | 2 | 120 |
13649 | G-100 | 3 | 230 |
13653 | G-350 | 4 and 5 | 880 |
*Certified value is printed on container label. Note: A graph is furnished with each cup showing efflux time of the "G" oil from 20 to 27 degrees Celsius. These standard oils prepared expressly by the Cannon Instrument Company for the Paul N. Gardner Company are produced in accordance with ISO/IEC 17025:2005, ISO/IEC Guide 34:2009, ISO 9001:2008. Caution: Silicone fluids should not be used to calibrate viscosity cups. These materials change the interface between the cup surface and the test material and therefore change the cup calibration. The following is taken from ASTM D445: Viscometers used for silicone fluids should be reserved for the exclusive use of such fluids. Solvent washings from these viscometers should not be used for cleaning other viscometers. |
Standards
Procedure
Downloads
Manuals
Knowledge
Want to learn the perfect process to using a viscosity cup? Find more information here about how to choose the right cup, how to measure efflux time and the best way to remove calibrations? CLICK HERE
Accessories
Other Variations